The affected device, evita v600, was examined onsite by dräger service and a faulty circuit board pba m48.3 was identified as root cause of the device malfunction.Die affected circuit board pba m48.3 was made available for further investigation at the manufacturer¿s site.The hardware analysis of the returned pba m48.3 circuit board revealed an overload at the input for the 5v voltage supply due to a short circuit of the ceramic capacitor c5.As a result, there was a complete loss of function of the pba m48.3 circuit board and the central control, ventilation and display functions of the device.The cause of the short circuit of the ceramic capacitor c5 could not be determined.Replacing the circuit board pba m48.3 remedied the issue, and the device was finally successfully tested according to the manufacturer's specifications.In the event of a complete loss of ventilator function, the safety valve automatically opens against ambient, allowing the patient to breathe spontaneously.The auxiliary audible alarm is activated to alert the user of the failure.This alarm is supplied from an independent power source and lasts for at least 2 minutes.In the current event, the device reacted as specified on the complete loss of device function and generated the auxiliary acoustic alarm as specified.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|