This incident has been recorded under cmp-(b)(4).Updates performed on b4, b5, d4, d9, g3, g6, h1, h2, h3, h4, h6, h10.Review of the most recent repair record determined the device was cutting skin irregularly.The control bar was out of position.The semi-circle bearings, vespel bearings, fine adjustment cams, and set screws were replaced and the control bar was repositioned correctly and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|