Model Number N/A |
Device Problems
Electrical /Electronic Property Problem (1198); Material Frayed (1262); Mechanical Problem (1384)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/05/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the device was not working and there was cut in the power cord.This was noted prior to surgery no harm or delay were reported as it relates to this event.No other adverse events were reported along with the event.
|
|
Manufacturer Narrative
|
This incident was submitted under (b)(4).Once investigation is completed a supplemental/final report will be submitted.
|
|
Event Description
|
No additional information is available.
|
|
Manufacturer Narrative
|
This incident has been recorded under (b)(6).Updates performed on b4 b5 d4 d9 g3 g6 h2 h3 h4 h6 h10 review of the most recent repair record determined the motor was corroded, the motor speed was unstable and the insulation of the cable was damaged.The motor and plug harness were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|