• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON FEMORAL DISTAL AUGMENT 5MM - SIZE 5 RIGHT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON FEMORAL DISTAL AUGMENT 5MM - SIZE 5 RIGHT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5540-A-502
Device Problem Loss of Osseointegration (2408)
Patient Problems Pain (1994); Inadequate Osseointegration (2646)
Event Date 05/24/2022
Event Type  Injury  
Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.
 
Event Description
It was reported that the patient's right knee was revised.As reported by rep: "pt.Presented with thigh pain, imaging was indicative of a loose femoral implant following a knee-revision "back in (b)(6) of 2017".Surgeon ruled out infection and determined the femoral construct was aseptically loose.He performed a revision of the femoral implants, leaving the tibial baseplate and stem in-situ.No further information available." spoke to rep.No wear or damage was noted intra-operatively.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON FEMORAL DISTAL AUGMENT 5MM - SIZE 5 RIGHT
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key14723112
MDR Text Key294256519
Report Number0002249697-2022-00856
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327015096
UDI-Public07613327015096
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K070095
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date05/31/2021
Device Model Number5540-A-502
Device Catalogue Number5540-A-502
Device Lot NumberVOUR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/24/2022
Initial Date FDA Received06/16/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age72 YR
Patient SexMale
-
-