The catalog number identified has not been cleared in the us but is similar to the fluency plus endovascular stent graft that are cleared in the us.The pro code and 510 k number for the fluency plus endovascular stent graft are identified.Manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Based on the information available it is not reasonably suggested that a manufacturing process may have caused or contributed to the reported issue.Investigation summary: the physical sample was not returned for evaluation.Pictures show a contaminated and partially deployed device.Based on the provided information the investigation is confirmed for partial deployment.The reported use of the device to treat an aneurysm represents an off-label use.A definite root cause for the reported event could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling, it was found that the instructions for use sufficiently address the potential risks.Regarding preparation of the device the instructions for use state that 'prior to loading the vascular system over a guide wire, both ports must be flushed with sterile saline.Flushing these lumens will also facilitate stent graft deployment.' regarding the anatomy of the placement site the instructions for use states: 'prior to stent graft deployment, ensure that the proximal stent graft end is positioned in a straight section of the lumen to reduce the risk of increased deployment forces and possible failure to deploy.' regarding accessories the instructions for use states: 'a super stiff guide wire (0.035 in.) is advanced from a femoral artery puncture site.Use an introducer sheath for the implant procedure'; the packaging pictograms indicate an introducer size of 10f and a 0.035" guidewire.It also contains the following precaution: 'the safety and effectiveness of the device for use in the treatment of aneurysms and pseudoaneurysms have not been established'.(expiry date: 03/2025).Device not returned.
|