• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION SOLUTION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Catalog Number ABM9830L
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
First name: (b)(6).Initial reporter address: initial reporter postal code: should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the packaging of an unspecified quantity of solution sets were damaged (torn).The issue was identified before patient use.There was no patient involvement.No additional infomraiton is available.
 
Manufacturer Narrative
Additional information was added to d9, h3, h4, h6 and h10.H10: the device was received for evaluation.Visual inspection was performed which observed a damaged on the packaging.The reported condition was verified.The cause of the condition could not be determined.Meanwhile, ten (10) retained samples were visually inspected and no issues were found.The reported condition was not verified in the retained samples.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOLUTION SET
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ATLACOMULCO
av. ing. salvador sanchez
colin no. 9
atlacomulco CP 50 450
MX   CP 50450
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key14735401
MDR Text Key294674551
Report Number1416980-2022-03176
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00085412353234
UDI-Public(01)00085412353234
Combination Product (y/n)N
Reporter Country CodeMX
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberABM9830L
Device Lot Number56CLE
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 05/20/2022
Initial Date FDA Received06/17/2022
Supplement Dates Manufacturer Received06/30/2022
Supplement Dates FDA Received07/08/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/04/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-