• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X42MM; DELTA XTEND IMPLANTS : SCREWS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US DXTEND SCREW LOCK D4.5X42MM; DELTA XTEND IMPLANTS : SCREWS Back to Search Results
Model Number 130790042
Device Problem Osseointegration Problem (3003)
Patient Problems Unspecified Infection (1930); Inadequate Osseointegration (2646)
Event Date 06/06/2022
Event Type  Injury  
Event Description
It was reported that the patient had loosening of the metaglene due to infection.Cement manufacturer was unknown.There was no surgical delay.Doi: unknown.Dor: (b)(6) 2022.Affected side: unknown shoulder.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : an analysis of the product could not be performed since a physical sample was not received for evaluation.An evaluation of the manufacturing record could not be performed as the required product identification number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.However, if the product or product id numbers are received at a later date, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND SCREW LOCK D4.5X42MM
Type of Device
DELTA XTEND IMPLANTS : SCREWS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY INT'L LTD. 8010379
st anthonys road
leeds LS11 8 DT
UK   LS11 8 DT
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key14743264
MDR Text Key294494969
Report Number1818910-2022-11240
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027867
UDI-Public10603295027867
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number130790042
Device Catalogue Number130790042
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
DXTEND SCREW LOCK D4.5X24MM.; DXTEND SCREW LOCK D4.5X42MM.; GLENOSPHERE ECCENTRIC +0 42MM.; GLOBAL UNITE SZ 12 STANDARD STEM.; HUMERAL PE CUP SZ 42 +3.; METAGLENE.; SZ 1 CENTERED EPIPHYSIS.
Patient Outcome(s) Required Intervention;
-
-