ZIMMER BIOMET, INC. CONSTRAINED TIBIAL ARTICULAR SURFACE PROVISIONAL SHIM SIZE J 10 MM THICKNESS; PROSTHESIS, KNEE
|
Back to Search Results |
|
Model Number N/A |
Device Problems
Detachment of Device or Device Component (2907); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that a tasp shim was worn and missing bearings.Attempts to obtain additional information have been made; however, no more information is available.
|
|
Manufacturer Narrative
|
(b)(4).Visual examination of the returned product found that it exhibits signs of repeated use (nicked or gouged) and has both of components one and two disassembled / missing.The missing components were not returned.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.The device is confirmed to have been a part of a capa investigation.Field actions were initiated to recommend against using ultrasonic cleaning as a sterilizing technique for these devices.These devices were subsequently re-designed to account for this failure mode with a new locking mechanism.It was determined that these devices were manufactured prior to this design change.The root cause is considered to be a previously addressed design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|