• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. REMSTAR PRO C-FLEX+; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. REMSTAR PRO C-FLEX+; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DS460HS
Device Problem Degraded (1153)
Patient Problems Angina (1710); Headache (1880); Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 09/23/2021
Event Type  malfunction  
Event Description
The manufacturer received information alleging an issue related to a cpap device's sound abatement foam.Patient alleged visualization of particles in the tubing chamber also alleged headache and chest pressure.There was no report of patients serious or permanent harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously received information alleging issue related to a cpap device's sound abatement foam.Section b5 has been corrected and should be reported as: the manufacturer received information alleging headches, upper airway irritation, chest pressure, black particles in the tubing/partciles in device related to a cpap device's sound abatement foam.There was no report of patient harm or injury.Repeated attempts to have the device and components returned for evaluation and investigation were unsuccessful.The manufacturer believes they will be unable to gather additional information.The manufacturer is submitting a final report at this time.If pertinent information becomes available to the manufacturer at a later date, an addendum to this final report will be filed.Section h6 corrected (health effect clinical code) and updated in this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR PRO C-FLEX+
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14748356
MDR Text Key301944378
Report Number2518422-2022-43593
Device Sequence Number1
Product Code BZD
UDI-Device Identifier00606959005839
UDI-Public00606959005839
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/20/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDS460HS
Device Catalogue NumberDS460HS
Was Device Available for Evaluation? No
Date Manufacturer Received10/25/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
-
-