Model Number 27081
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/23/2022
Event Type  malfunction  
Manufacturer Narrative
The astral device was returned to a resmed authorized third party service centre.Evaluation confirmed the reported complaint.The pneumatic block assembly was replaced to address the issue.The device was serviced and fully tested before it was returned to the customer.Resmed reference#: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf188) indicating a safety assert system fault.There was no patient harm or serious injury reported as a result of this incident.
 
Manufacturer Narrative
The astral device was returned to resmed for an investigation.Performance testing confirmed the reported complaint.Based on all available evidence and complaint investigations of a similar nature, the investigation determined that the reported complaint was due to contamination.Resmed¿s risk analysis for this failure mode concludes that the risk is acceptable.Resmed reference: (b)(4).
 
Event Description
It was reported to resmed that an astral device displayed an error message (sf188) indicating a safety assert system fault.There was no patient harm or serious injury reported as a result of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASTRAL 100 - APAC1
Type of Device
Ventilator, continuous, facility use
Manufacturer (Section D)
RESMED LTD
1 elizabeth macarthur drive
bella vista
sydney, nsw 2153
AS  2153
MDR Report Key14758325
MDR Text Key295249798
Report Number3007573469-2022-00429
Device Sequence Number1
Product Code CBK
UDI-Device Identifier00619498270811
UDI-Public(01)00619498270811(10)1440081
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation 505
Type of Report Initial,Followup
Report Date 12/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number27081
Device Catalogue Number27081
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/30/2022
Was the Report Sent to FDA? Yes
Date Report Sent to FDA12/28/2022
Distributor Facility Aware Date11/30/2022
Device Age26 MO
Date Report to Manufacturer12/28/2022
Initial Date Manufacturer Received Not provided
Initial Date FDA Received06/21/2022
Supplement Dates Manufacturer Received11/30/2022
Supplement Dates FDA Received12/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-