Model Number CDDRA500Q
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2022
Event Type  Injury  
Event Description
It was reported that the patient presented remotely via merlin.Net.Review of transmission revealed charge time out on the implantable cardioverter defibrillator (icd).No changes or intervention was reported.The patient condition was stable.
 
Event Description
New information indicates that the implantable cardioverter defibrillator (icd) was explanted and replaced on (b)(6) 2022.The patient condition was stable.
 
Manufacturer Narrative
The reported field event of charge time out was confirmed.Review of device data showed the device had a charge time out during capacitor maintenance.The charge time out was due to an internal anomaly in the high voltage capacitor.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT DR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14781658
MDR Text Key295895701
Report Number2017865-2022-13154
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032003
UDI-Public05415067032003
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 09/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2021
Device Model NumberCDDRA500Q
Device Catalogue NumberCDDRA500Q
Device Lot NumberS000077570
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/26/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/28/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
2088TC/52 TENDRILLEADS; 7122Q/65 DURATATACHY
Patient Outcome(s) Required Intervention;
Patient Age82 YR
Patient SexMale
-
-