Model Number 560BCS1 |
Device Problems
Display or Visual Feedback Problem (1184); Smoking (1585)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/09/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a bio-console 560 instrument, it was reported that there was a burnt smell and smoke and a black screen was seen when booting.Use of the instrument was unknown.There were no adverse patient effects because of this issue.Medtronic received additional information stating, that information on whether the hand crank was used was not provided.The bio-console was being used for an extra-corporal membrane oxygenation (ecmo) case.
|
|
Manufacturer Narrative
|
Product analysis: the reported burnt smell and smoke and a black screen seen when booting was verified during service.The issue was resolved by replacing the assy 560ui bitsy xb 2u02011 led display, pcba 560 display backlight pwm and sub-assy ui 560 bio-console.The assy 560 bio-console syscon was replaced with a new version of the part.The instrument passed all testing.Preventive maintenance was performed per specifications.Conclusion: the complaint of the reported burnt smell and smoke and a black screen seen when booting was verified during service.This occurrence is the result of the replacement of a component used in the previous version of the assy 560 bio-console syscon.There were no patient/clinical safety issues reported.Trends for issues with this instrument are monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|