W. L. GORE & ASSOCIATES, INC. GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS; ILIAC COVERED STENT, ARTERIAL Back to Search Results
Model Number BXA097902A
Device Problems Obstruction of Flow (2423); Deformation Due to Compressive Stress (2889)
Patient Problem Obstruction/Occlusion (2422)
Event Date 09/10/2021
Event Type  Injury  
Manufacturer Narrative
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.Due to no device return, an investigation could not be performed.Analysis of production records will be provided in final report.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event Description
Subject: 1701-181-004, onset date: (b)(6) 2021.Description: compression of the superior mesenteric artery during procedure.Outcome: recovered/resolved - treatment required: angio demonstrated some compression of the superior mesenteric artery.Recatherized the superior mesentric artery and ballooned the stent graft at the compressed segment.Flush aortagram demonstrated an excellent result throughout with brisk flow.
 
Manufacturer Narrative
A review of the manufacturing records indicated the lot met pre-release manufacturing specifications.
 
Manufacturer Narrative
Added suspected product to section c.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS
Type of Device
ILIAC COVERED STENT, ARTERIAL
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL ECHO RIDGE B/P
3250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
spencer deboard
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key14788840
MDR Text Key294859146
Report Number2017233-2022-03038
Device Sequence Number1
Product Code PRL
UDI-Device Identifier00733132637607
UDI-Public00733132637607
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P160021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/24/2023
Device Model NumberBXA097902A
Device Catalogue NumberBXA097902A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/01/2021
Initial Date FDA Received06/23/2022
Supplement Dates Manufacturer Received12/01/2021
12/01/2021
Supplement Dates FDA Received07/13/2022
09/19/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2020
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age67 YR
Patient SexMale
Patient Weight77 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-