Model Number PE074F5
Device Problem Pacing Problem (1439)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/03/2022
Event Type  malfunction  
Event Description
It was reported a swan ganz was unable to pace from the beginning after the catheter insertion during urgent catheterization.The issue was resolved by replacing the catheter.It is unknown if the patient had cardiac conduction defect.Patient demographic information was requested but unavailable.The device was discarded at the hospital.There were no patient complications reported.Customer was unsure of lot number, but provided four possible numbers, 64206873 or 64113324 or 64113325 or 64069136.
 
Manufacturer Narrative
The device was not returned for evaluation.Device was discarded at hospital.Without return of the unit it is not possible to determine if some damage or defect existed on the unit that could have contributed to the event.It is not known if some procedural factors may have contributed to the event.Further evaluation regarding supplier related quality issues is under investigation.A device history record review for the 4 possible lot numbers was completed and all four reports documented that device met all specifications upon distribution.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRANSLUMINAL BIPOLAR PACING PROBE
Type of Device
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 94614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
jessica atallah
1 edwards way
irvine, CA 92614
MDR Report Key14794332
MDR Text Key301940778
Report Number2015691-2022-06354
Device Sequence Number1
Product Code LDF
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K813521
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date03/06/2024
Device Model NumberPE074F5
Device Lot Number64206873
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/06/2022
Initial Date FDA Received06/23/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/10/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-