MAUDE Adverse Event Report: ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING Back to Search Results
Model Number D-AVHD-DF16
Device Problem Signal Artifact/Noise (1036)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/31/2022
Event Type  malfunction  
Event Description
Manufacturer ref# 3008452825-2022-00460.During the supraventricular tachycardia procedure, noise was noted on the catheter which caused a delay.The connecting cable, pin block and catheter were exchanged, with no resolution.The catheter was exchanged again and the procedure was completed with no adverse patient consequences.
 
Manufacturer Narrative
One bi-directional, curve d-f, sensor enabled, advisor hd grid mapping catheter was received for evaluation.Electrodes 1-18 met specifications for acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported noise and subsequent delay remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14795912
MDR Text Key301663990
Report Number3008452825-2022-00461
Device Sequence Number1
Product Code DRF
UDI-Device Identifier05415067028198
UDI-Public05415067028198
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172393
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberD-AVHD-DF16
Device Catalogue NumberD-AVHD-DF16
Device Lot Number8318048
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/11/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿
-
-