• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VDW GMBH WST 6:1 F. VDW.GOLD/SILVER; HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL Back to Search Results
Catalog Number V041079000000
Device Problems Break (1069); Mechanical Problem (1384); Therapeutic or Diagnostic Output Failure (3023)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
In this event it is reported that wst 6:1 f.Vdw.Gold/silver contra angle does not rotate properly, unbalanced.And involved in file breakages.Files are unknown as of this mdr.No injury occurred.Device and contra angle will be returned for investigation.
 
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Manufacturer Narrative
After a thorough check, we could not find any error.Foot control 202214, gmr2294074 and contra angle155518 (2022) without errors.Torque test carried out.Vdw gold reciproc with serial number gr60471 meets specification concerning the torque limitation function.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WST 6:1 F. VDW.GOLD/SILVER
Type of Device
HANDPIECE, CONTRA- AND RIGHT-ANGLE ATTACHMENT, DENTAL
Manufacturer (Section D)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM  81737
Manufacturer (Section G)
VDW GMBH
bayerwaldstrasse 15
munich bayern 81737
GM   81737
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key14800657
MDR Text Key301115596
Report Number9611053-2022-00574
Device Sequence Number1
Product Code EGS
Combination Product (y/n)N
Reporter Country CodeGM
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 09/02/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberV041079000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date06/23/2022
Initial Date Manufacturer Received 06/23/2022
Initial Date FDA Received06/24/2022
Supplement Dates Manufacturer Received06/23/2022
Supplement Dates FDA Received09/02/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-