Catalog Number V041079000000 |
Device Problems
Break (1069); Mechanical Problem (1384); Therapeutic or Diagnostic Output Failure (3023)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
In this event it is reported that wst 6:1 f.Vdw.Gold/silver contra angle does not rotate properly, unbalanced.And involved in file breakages.Files are unknown as of this mdr.No injury occurred.Device and contra angle will be returned for investigation.
|
|
Manufacturer Narrative
|
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
|
|
Manufacturer Narrative
|
After a thorough check, we could not find any error.Foot control 202214, gmr2294074 and contra angle155518 (2022) without errors.Torque test carried out.Vdw gold reciproc with serial number gr60471 meets specification concerning the torque limitation function.
|
|