A us distributor contacted zoll to report that a patient passed away on (b)(6) 2022 while reportedly wearing the lifevest.The patient received an appropriate treatment which converted the arrhythmia to a slower rhythm.The patient also received a non-lifevest defibrillation. the device was started up at 09:03:37 on (b)(6) 2022.At 12:18:56, an arrhythmia was detected.Ecg shows af with rvr @ 150 bpm with motion artifact.The rhythm then degrades to vt @ 180 bpm.At 12:19:32, the patient received the appropriate treatment.Rhythm at time of treatment was vt @ 150 bpm with motion artifact.Post shock rhythm was sinus bradycardia from 30-10 bpm with motion artifact and hb.The rhythm then transitions to sinus bradycardia @ 40 bpm with hb.At 12:37:10, an arrhythmia was detected.Ecg shows idioventricular rhythm @ 50 bpm degrading to vf with varying amplitudes.The rhythm then degrades to asystole.The device properly detected vf.However, varying amplitude and asystole prevented the lifevest from treating the patient.From 12:58:48 to 13:12:44, an arrhythmia was detected four times.Ecg shows asystole with cpr/motion artifact/electrode lead fall off and cardiac activity.The rhythm then degrades to vf with cpr/motion artifact and electrode lead fall off.The device properly detected vf.However, cpr/motion artifact/electrode lead fall off prevented the lifevest from treating the patient.At 13:12:44, the patient received a non-lifevest defibrillation.Rhythm at time of treatment was vf with cpr/motion artifact and electrode lead fall off.Post shock rhythm was asystole with cpr/motion artifact and electrode lead fall off.The electrode belt was disconnected at 13:15:42 on (b)(6) 2022.The patient passed away on (b)(6) 2022.
|