The initial report was submitted under manufacturer report number 3008344661-2022-00066-01 abbott (b)(4) as manufacturing site)with suspect medical device of architect cyclosporine reagent kit, list 3r30.After further evaluation, the suspect medical device was changed to architect cyclosporine whole blood precipitation reagent kit, list 1l75 abbott (b)(4) as manufacturing site , which is under this report.
|
This follow-up submission being send to cross reference mdr 1415939-2022-00060-00 for likely cause changed on 21jul2022; to architect cyclosporine reagent kit, list number 01l75-25, lot number 37701fp00 from architect cyclosporine whole blood precipitation reagent kit, list 01l75-55, lot number 310275.After further evaluation it was determined that the cause of the issue was the architect cyclosporine reagent kit, list 01l75-25, lot number 37701fp00.) (abbottpark, il as manufacturing site) and submitted under manufacturer report number 1415939-2022-00060-00.
|