Related manufacturer reference number:2017865-2022-13453, related manufacturer reference number:2017865-2022-13454.It was reported that the right ventricle lead was dislodged, identified via fluoroscopy and also exhibited failure to capture.The right ventricle lead was repositioned on (b)(6) 2022.Patient was stable.On (b)(6) 2022, it was noted that the left ventricle lead and right atrial lead were both dislodged, identified via fluoroscopy, and both the lead exhibited no pacing and extra cardiac stimulation.Both that atrial and left ventricle lead were explanted and replaced.Patient was stable.
|
The reported events were lead dislodgement, no pacing, and therapy delivered to incorrect body area.As received, a complete lead was returned in one piece.The reported event of no pacing was not confirmed.Visual and x-ray examination of the lead did not find any anomalies.Electrical testing did not find any indication of conductor fractures or internal shorts.S-curve was measured to be within specification.
|