• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problems Computer Software Problem (1112); Insufficient Information (3190)
Patient Problems Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2021
Event Type  malfunction  
Event Description
As customer described 'aps configuration error.Error comes on device when powered on.Philips rdt planning to receive the device and send to the manufacturer for the root cause investigation.
 
Manufacturer Narrative
Philips received a complaint on the tempus ls exhibiting aps configuration error.Error comes on device when powered on.The device sent to schiller for investigation.No device failure found during the entry test.The investigation on the log file showed that this error message was logged only once - not reproducible.During start-up the ecg microchip is configured with the system parameters.Only when the ecg microchip cannot be configured successfully, the device displays this error message.As the error is not reproducible sag cannot determine what triggered this error.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The investigation concludes that no further action is required at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key14854432
MDR Text Key302962116
Report Number3003832357-2022-00013
Device Sequence Number1
Product Code LDD
UDI-Device Identifier07613365002737
UDI-Public07613365002737
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/23/2021
Initial Date FDA Received06/29/2022
Supplement Dates Manufacturer Received11/23/2021
Supplement Dates FDA Received08/18/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/20/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-