Philips received a complaint on the tempus ls exhibiting aps configuration error.Error comes on device when powered on.The device sent to schiller for investigation.No device failure found during the entry test.The investigation on the log file showed that this error message was logged only once - not reproducible.During start-up the ecg microchip is configured with the system parameters.Only when the ecg microchip cannot be configured successfully, the device displays this error message.As the error is not reproducible sag cannot determine what triggered this error.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The investigation concludes that no further action is required at this time.
|