Catalog Number 42055150-120 |
Device Problem
Material Separation (1562)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 06/10/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
The device will be returned for analysis.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that the procedure was to treat a right superficial femoral artery.A 5.5x150mm supera device was advanced to the lesion and deployed without issue.The delivery system was removed outside of the patient when it was noted that the white taper tip was not present on the device as it was removed off the guide wire.Fluoroscopy demonstrated that the tip was still on the guide wire, but at the end of the 6fr sheath.A snare device was used to successfully remove the broken tip.A clinically significant delay was reported as intervention was performed to remove the broken tip.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that during removal interaction with the 6fr sheath resulted in the reported tip separation.The treatment(s) appears to be related to the operational context of the procedure as a snare device was used to successfully remove the broken tip.A clinically significant delay was reported as intervention was performed to remove the broken tip.There is no indication of a product quality issue with respect to manufacture, design or labeling.D9, h3: device status changed from yes to no.
|
|
Search Alerts/Recalls
|