Model Number N/A |
Device Problems
Failure to Cut (2587); Physical Resistance/Sticking (4012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/01/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that prior to surgery the device produced an irregular cut and had a stuck motor.No harm or delay were reported as it relates to this event.No adverse events were reported as it relates to this event.
|
|
Manufacturer Narrative
|
This incident has been submitted under (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This incident has been recorded under (b)(6).Updates performed on b4 b5 d4 d9 g3 g6 h2 h3 h4 h6 h10 review of the most recent repair record determined that the unit had cable contact issues and found possible corrosion on the motor.The motor and plug harness assembly were replaced and resolved the reported issue.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information is available.
|
|