BOSTON SCIENTIFIC CORPORATION AUTOGEN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number G179 |
Device Problems
High impedance (1291); Noise, Audible (3273)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/03/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.
|
|
Event Description
|
It was reported that the patient with this cardiac resynchronization therapy defibrillator (crt-d) heard beep tones over the last three to four days.The shock impedance measurement from right atrial (ra) to right ventricular (rv) is about 183 ohms, shock impedance measurement was at 188 ohms from ra to can.No noise was noted and this device was reprogrammed to rv to can.Boston scientific technical services suggested consideration to perform defibrillation threshold testing (dft) given the shock vector has changed.The patient has a follow-up appointment in july.No adverse patient effects were reported.This device and competitor lead remains in-service.
|
|