MEDTRONIC PUERTO RICO OPERATIONS CO. ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number 37612 |
Device Problem
Delayed Charge Time (2586)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/02/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient had been having trouble charging their implant because typically they would put the recharger antenna in their bra and start charging and fall asleep.Patient said in the last couple weeks the implant had been taking longer to charge because the recharger antenna would slip out and wouldn't stay in place.The patient confirmed they got good coupling while charging.Additional information was received from the consumer who reported that the adhesive discs helped them, but their implant was still taking longer to charge up.The patient said they didn't let their implant go below 75% and it would normally take about an hour to charge from 75-100% but now it was taking 1 hour and 50 minutes.Patient said they got full coupling.The patient didn't have recharger with them but based off information, patient services (ps) sent replacement recharger antenna.The patient mentioned that they went from charging once a day to charging twice a day, ps tried to confirm if patient was saying that the implant battery was de pleting more quickly than before, patient went back and forth but then said the battery wasn't depleting differently and requested replacement recharger antenna be sent.Patient said there had been no falls/trauma/change to settings or implant site.Additional information was received from the consumer who reported that they received and are using the replacement recharger antenna but issue persisted.They further described that they charge up their ins every day.The caller stated they charged their ins fully yesterday, which took an hour and a half, which was longer than their normal, and then charged again today for an hour.The caller confirmed they have not had any noticeable changes to their implant site, and have not had any recent weight loss/gain.Redirected patient to their hcp to further address the issue.
|
|
Search Alerts/Recalls
|
|
|