Model Number MOSAIQ
Device Problem Computer Software Problem (1112)
Patient Problems Unintended Radiation Exposure (4565); Insufficient Information (4580)
Event Date 05/23/2022
Event Type  malfunction  
Event Description
The customer reported wrong field import.
 
Manufacturer Narrative
The manufacturer's investigation is on-going and further information will be provided once the investigation has completed.
 
Manufacturer Narrative
The investigation was completed by conducting a thorough evaluation of the product and the reported information.During in-house testing, elekta were unable to reproduce the issue when importing the customer's plans.The patient was treated with the wrong energy, collimator angle and jaw size for six of the ten treatments.If this is unaccounted for this would result in an underdose for this patient which is considered non-serious.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIQ
Type of Device
ACCELERATOR, LINEAR, MEDICAL
Manufacturer (Section D)
ELEKTA SOLUTIONS AB
kungstensgatan 18
box 7593
stockholm, SE-10 3 93
SW  SE-103 93
Manufacturer (Section G)
ELEKTA SOLUTIONS AB
400 perimeter center terrace
suite 50
atlanta GA 30346
Manufacturer Contact
pms
cornerstone
london road
crawley, west sussex RH10 -9BL
UK   RH10 9BL
MDR Report Key14884107
MDR Text Key295240651
Report Number3015232217-2022-00005
Device Sequence Number1
Product Code IYE
UDI-Device Identifier00858164002091
UDI-Public(01)00858164002091(10)2.60.342
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K141572
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberMOSAIQ
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/03/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-