OLYMPUS MEDICAL SYSTEMS CORP. EVIS EXERA DUODENOVIDEOSCOPE Back to Search Results
Model Number TJF-160VR
Device Problems Material Frayed (1262); Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/01/2022
Event Type  malfunction  
Manufacturer Narrative
The olympus fst was on site at the facility and repaired the device.An endocheck was performed, opened/closed control part, replaced albaran cable and soldered/adjusted stopper, replaced necessary small parts and o-rings, and function tested/leak tested after repair.According to the fst, the defect could have resulted in an injury to the mucous membrane.The spliced albaran cable was a known phenomenon, due to design.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.A supplemental report will be submitted upon completion of the investigation.
 
Event Description
The olympus representative reported the albaran bowden cable spliced on the evis exera duodenovideoscope.The issue was detected by the field service technician (fst) during inspection service.The user was not aware of the problem.No patient harm reported.During the evaluation of the device, it was noted some of the wires that compose the forceps elevator wire were raised due to cut or fray.This report is to capture the reportable malfunction of some of the wires that compose the forceps elevator wire were raised due to cut or fray noted at estimation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 14 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was presumed to have been due to a defective wire (designated as the "k-wire").It was not possible to determine or surmise further the cause of the defect.Users can detect the event in accordance with the following instructions for use (ifu): "be sure to prepare another endoscope to avoid that the examination will be interrupted due to equipment failure or malfunction." olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA DUODENOVIDEOSCOPE
Type of Device
DUODENOVIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14885119
MDR Text Key303669941
Report Number8010047-2022-11169
Device Sequence Number1
Product Code FDT
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K024033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTJF-160VR
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/01/2022
Initial Date FDA Received07/01/2022
Supplement Dates Manufacturer Received09/15/2022
Supplement Dates FDA Received10/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-