• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MONUMENT DEPUY SYNTHES PRODUCTS INC FIBULINK® SYNDESMOSIS REPAIR KIT/SS; WASHER, BOLT NUT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MONUMENT DEPUY SYNTHES PRODUCTS INC FIBULINK® SYNDESMOSIS REPAIR KIT/SS; WASHER, BOLT NUT Back to Search Results
Model Number FGS-1000
Device Problem Break (1069)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 01/01/2022
Event Type  Injury  
Event Description
Device report from synthes reports an event in the usa as follows: it was reported that on an unknown date, fibulink® syndesmosis repair kit/ss was broken.This complaint involves one (1) device fibulink® syndesmosis repair kit/ss.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from photos.Visual analysis of the photos revealed that the fibulink® syndesmosis repair kit/ss was broken.X-ray evidence "(b)(4) x-rays (2)" shows the fibula link being migrated from the fibula tensioning cap.In evidence "(b)(4) x-rays (1)" the fibula link appears to have broken from the left end, fragment remained inside the fibula tensioning cap.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the fibulink® syndesmosis repair kit/ss.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBULINK® SYNDESMOSIS REPAIR KIT/SS
Type of Device
WASHER, BOLT NUT
Manufacturer (Section D)
MONUMENT DEPUY SYNTHES PRODUCTS INC
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wright lane east
west chester, PA 19380
3035526892
MDR Report Key14914150
MDR Text Key295226466
Report Number1719045-2022-00021
Device Sequence Number1
Product Code HTN
UDI-Device Identifier00863176000300
UDI-Public(01)00863176000300
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162805
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFGS-1000
Device Catalogue NumberFGS-1000-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-