Product complaint # (b)(4).Additional narrative: complainant part is not expected to be returned for manufacturer review/investigation.(b)(4).Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from photos.Visual analysis of the photos revealed that the fibulink® syndesmosis repair kit/ss was broken.X-ray evidence "(b)(4) x-rays (2)" shows the fibula link being migrated from the fibula tensioning cap.In evidence "(b)(4) x-rays (1)" the fibula link appears to have broken from the left end, fragment remained inside the fibula tensioning cap.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the fibulink® syndesmosis repair kit/ss.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|