Model Number 4FC12
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2022
Event Type  malfunction  
Event Description
After a completed case, the sheath subsequently tested out of specification per the manufacturer's investigation.
 
Manufacturer Narrative
The 4fc12 sheath with lot number 0011045533 was returned and analyzed.Visual inspection before functional testing and dissection was performed on the shaft, handle, and dilator.No anomaly was identified during the external visual inspection.The handle, shaft and sideport were intact with no apparent issue.Insertion of the dilator into the sheath was performed.Unable to lock the dilator luer to sheath.Further inspection under microscope identified dilator luer damage, which may have caused the dilator to have difficulties locking with the sheath.In conclusion, the sheath failed the returned product inspection due to the damaged dilator luer.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key14917159
MDR Text Key303234786
Report Number3002648230-2022-00261
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/09/2024
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number0011045533
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2022
Initial Date Manufacturer Received 06/21/2022
Initial Date FDA Received07/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-