U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX700H11C
Device Problem Degraded (1153)
Patient Problems Sore Throat (2396); Taste Disorder (4422); Swelling/ Edema (4577)
Event Date 06/30/2022
Event Type  malfunction  
Event Description
The manufacturer received information alleging an issue related to a bipap device's sound abatement foam.The patient has alleged to experiencing a swollen/irritated/sore throat, spitting up reddish/pink fluid and the device has a strange taste.There was no report of patient harm or injury.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Manufacturer Narrative
The manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an allegation of an issue related to sound abatement foam became degraded and caused swollen/irritated/sore throat, spitting up reddish/pink fluid and the device has a strange taste.This event is assessed as related to the device in this case.Based on the available information, the manufacture concludes no further action is necessary.There was no medical intervention required by the patient.The device was returned to the manufacturer's service center for further evaluation.The device was evaluated.There was no mention of visual findings to the external part of the device.The internal aspect of the device was inspected.The device powered on and airflow was confirmed.The device's downloaded logs were reviewed by the manufacturer.There were no errors found.The manufacturer concludes that they could not confirm the customer's allegation and there was no visible foam degradation.
 
Manufacturer Narrative
Section b5 in previous report should be reported as : the manufacturer was contacted in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.The manufacturer previously reported an allegation of an issue related to sound abatement foam became degraded and caused swollen/irritated/sore throat, spitting up reddish/pink fluid and the device has a strange taste.There was no report of serious patient harm or injury.After review, the following correction has been updated in this report.Section(s) b1, b2 has updated to reflect as product problem.Section h1 has changed to reflect a malfunction.Section h6 health effect- impact code has been updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15206
2673970028
MDR Report Key14917188
MDR Text Key295259412
Report Number2518422-2022-58556
Device Sequence Number1
Product Code BZD
UDI-Device Identifier00606959025691
UDI-Public00606959025691
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 11/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/05/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX700H11C
Device Catalogue NumberDSX700H11C
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received10/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/09/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberRES 88058
Patient Sequence Number1
Patient Outcome(s) Other;
-
-