The distributor reported on behalf of the customer that the device, 410-2000 cga, surefit ground pad with 10ft cabel, 100/case was being used on an unknown date and " we have been having problems with the conmed bovie grounding pads that we are getting in our packs.They are not sticking well to most patients and causing issues mid case¿.There was no impact or injury to the patient or user.There was no report of medical intervention or any prolonged hospitalization for the patient.Further assessment found that no other information was known.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|
The device is not being returned and no photographic evidence was provided.Therefore, a device malfunction cannot be verified.A dhr review cannot not be conducted as no lot number was provided.A two-year lot history review cannot be conducted as no lot number was provided.A two-year review of complaint history revealed there has been a total of 53 complaints, regarding 492 devices, for this device family and failure mode.During this same time frame 10,373,340 devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the ifu, the user is advised to not attempt to select a well vascularized site in close proximity to the surgical site (anterior arm or thigh is recommended).Avoid placement on bony prominence, skin lesions or folds, tattoos, scars, metal prosthesis or near ecg electrodes and cables.Do not apply where fluid might pool.Always use the largest size pad per patient weight guidelines which can be properly applied to the site: adult products are for use on patients weighing more than 15 kg.Pediatric pads are for use on patients weighing between 5 and 15 kg.Prepare the skin at the application site according to facility protocol.If no protocol exists, clip excessive hair at application site, clean and disinfect area to remove oils, lotions, etc., and allow to dry.Do not open package until ready to apply to skin.Inspect the pad and cable.Do not use if product is expired or apparently damaged.Check expiration date on package.This issue will continue to be monitored through the complaint system to assure patient safety.
|
The distributor reported on behalf of the customer that the device, 410-2000 cga, surefit ground pad with 10ft cabel, 100/case was being used on an unknown date and " we have been having problems with the conmed bovie grounding pads that we are getting in our packs.They are not sticking well to most patients and causing issues mid case¿.There was no impact or injury to the patient or user.There was no report of medical intervention or any prolonged hospitalization for the patient.Further assessment found that no other information was known.This report is being raised on the basis of malfunction with potential for injury upon reoccurrence.
|