• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM MEXICO ADVANTA 2 FRAME; BED, MANUAL Back to Search Results
Model Number P1190A000067
Device Problem Unintended Movement (3026)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/09/2022
Event Type  malfunction  
Event Description
Hillrom received a report from a hillrom technician stating the bed was going up and down on its own.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint # (b)(4).
 
Manufacturer Narrative
The hillrom technician found the patient left caregiver control needed to be replaced.Per the hillrom service manual the advanta¿ 2 should be subject to an effective maintenance program.Test each of the buttons to check that they activate the correct function and that they do not work intermittently by pressing each button for several seconds.Each movement must be continuous.Replace the pendant if necessary.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed in (b)(6) 2022.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the patient left caregiver control to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA 2 FRAME
Type of Device
BED, MANUAL
Manufacturer (Section D)
HILL-ROM MEXICO
ave.del telefono no. 200 col. huinala
apodaca, nuevo leon 66640
MX  66640
Manufacturer Contact
brad wheeler
1069 state route 46 east
batesville, IN 47006
3128199307
MDR Report Key14919177
MDR Text Key302717167
Report Number3006697241-2022-00043
Device Sequence Number1
Product Code FNJ
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 07/05/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberP1190A000067
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/09/2022
Initial Date FDA Received07/05/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/17/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-