• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET UK LTD. OXF TWIN PEG CMNTLS FMRL MD; KNEE PROTHESIS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET UK LTD. OXF TWIN PEG CMNTLS FMRL MD; KNEE PROTHESIS Back to Search Results
Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Pain (1994); Synovitis (2094)
Event Date 01/27/2022
Event Type  Injury  
Manufacturer Narrative
(b)(4).Concomitant medical devices: oxf anat brg rt mdsize 3 pma; item# 159575; lot# 816870; oxf uni cmntls tib szd rm; item# (b)(4); lot# r3050482a.Multiple mdr reports were filed for this event, please see associated reports: 3002806535-2022-00290, 3002806535-2022-00291, 3002806535-2022-00292.The device history records for were reviewed for deviations and / or anomalies and no discrepancies related to the reported event were found.Review of x-ray images (6week post-op and 3 year post-op) show overall fit, alignment, and bone quality appear normal.There are no signs of loosening, wear, radiolucency, or other contributing factors.Soft tissues are normal, no hardware or bone fractures present.There are no other concerns noted that would lead to the patient's complaint of pain/catching sensation.The root cause of the event was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that a partial knee arthroplasty surgery was performed and approximately 6 years later the patient is still experiencing intermittent pain and swelling.Due diligence is in progress for this complaint; to date, no additional information or product has been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OXF TWIN PEG CMNTLS FMRL MD
Type of Device
KNEE PROTHESIS
Manufacturer (Section D)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK  CF31 3XA
Manufacturer (Section G)
BIOMET UK LTD.
waterton industrial estates
bridgend CF31 3XA
UK   CF31 3XA
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key14926775
MDR Text Key295314807
Report Number3002806535-2022-00290
Device Sequence Number1
Product Code NRA
UDI-Device Identifier05019279442270
UDI-Public(01)05019279442270(17)251113(10)R3050584A
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Study,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue Number161474
Device Lot NumberR3050584A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/06/2022
Initial Date FDA Received07/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/13/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age67 YR
Patient Weight75 KG
-
-