Additional information added: h6 (component code, type of investigation, investigation findings, investigation conclusions).H9: z-2745-2015 through z-2747-2015.Results of investigation: it was reported that revision of the femoral head component was performed due to loosening.As of today, the device, used in treatment, has not been returned for evaluation and further information has not been supplied in relation to the case.A review of the historical complaints data for the femoral head was performed.One other complaint was identified involving this batch number.Other similar complaints were identified for the part number and the reported/related failure mode.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the product and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.No details have been supplied to enable an evaluation of the case from a clinical / medical perspective.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Additionally, specific factors known to contribute to the alleged fault cannot be provided due to the insufficient information.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Corrected data: h3, h6(health effect - clinical code, health effect - impact code).
|