Catalog Number 74121146
Device Problem Loss of Osseointegration (2408)
Patient Problems Failure of Implant (1924); Joint Laxity (4526)
Event Date 06/15/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a bhr surgery, the resurfacing femoral head 46mm loosened in the femur so it was revised on (b)(6) 2022 with a bhr dual mobility and a stem.Current health status of the patient is unknown.
 
Manufacturer Narrative
D4 and h4 were corrected.
 
Manufacturer Narrative
Section d4 was updated.
 
Manufacturer Narrative
Additional information added: h6 (component code, type of investigation, investigation findings, investigation conclusions).H9: z-2745-2015 through z-2747-2015.Results of investigation: it was reported that revision of the femoral head component was performed due to loosening.As of today, the device, used in treatment, has not been returned for evaluation and further information has not been supplied in relation to the case.A review of the historical complaints data for the femoral head was performed.One other complaint was identified involving this batch number.Other similar complaints were identified for the part number and the reported/related failure mode.However, as the device is no longer sold, no action is to be taken.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.Review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the product and similar complaint events was performed.Following the review, prior applicable escalation actions were identified and confirmed to reduce associated risks as far as possible.No further escalation actions are required.No details have been supplied to enable an evaluation of the case from a clinical / medical perspective.Based on the information provided we cannot confirm or further investigate the reported complaint, our investigation remains inconclusive, and a definitive root cause cannot be determined.Additionally, specific factors known to contribute to the alleged fault cannot be provided due to the insufficient information.If additional information becomes available in the future, this case will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.Corrected data: h3, h6(health effect - clinical code, health effect - impact code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESURFACING FEMORAL HEAD 46MM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK  CV31 3HL
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS LTD
aurora house, spa park
leamington spa warwickshire CV31 3HL
UK   CV31 3HL
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14933209
MDR Text Key295345916
Report Number3005975929-2022-00415
Device Sequence Number1
Product Code NXT
UDI-Device Identifier03596010502780
UDI-Public3596010502780
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P040033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 09/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2013
Device Catalogue Number74121146
Device Lot Number089778
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/19/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-2267-2018, Z-2268-2018
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexFemale
-
-