Model Number CI632
Device Problem Positioning Failure (1158)
Patient Problem Hearing Impairment (1881)
Event Date 06/03/2022
Event Type  Injury  
Manufacturer Narrative
This report is submitted on july 7, 2022.
 
Event Description
Per the clinic, the patient experienced poor performance with the device which was due to a tip fold-over of the electrode at the initial implantation.Reprogramming attempts were made; however, the issue could not be resolved.The device was explanted on (b)(6) 2022.The patient was re-implanted with a new device in the same procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUCLEUS CI632 PROFILE PLUS WITH SLIM MODIOLAR ELECTRODE
Type of Device
NUCLEUS 24 COCHLEAR IMPLANT SYSTEM
Manufacturer (Section D)
COCHLEAR LTD
1 university avenue
macquarie university, nsw 2109
AS  2109
MDR Report Key14940825
MDR Text Key295400570
Report Number6000034-2022-01988
Device Sequence Number1
Product Code MCM
UDI-Device Identifier09321502036580
UDI-Public(01)09321502036580(11)211109(17)231108
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/07/2022,06/08/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/06/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date11/08/2023
Device Model NumberCI632
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/22/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA07/07/2022
Distributor Facility Aware Date06/08/2022
Date Report to Manufacturer06/08/2022
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
Patient SexFemale
-
-