• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG COVERA PLUS VASCULAR COVERED STENT Back to Search Results
Catalog Number AASLE06100
Device Problems Loss of or Failure to Bond (1068); Break (1069); Positioning Failure (1158)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2022
Event Type  malfunction  
Event Description
It was reported that during a stent graft placement procedure, the pulley of the stent allegedly broke.It was further reported that the stent allegedly failed to deploy.There was no reported patient injury.
 
Manufacturer Narrative
The catalog number identified has not been cleared in the us but is similar to the covera vascular covered stent products that are cleared in the us.The pro code and 510 k number for the covera vascular covered stent products are identified.As the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a video was provided for review.The investigation of the reported event is currently underway.(expiry date: 11/2022).Device pending return.
 
Event Description
It was reported that during a stent graft placement procedure via common femoral puncture through contralateral approach, the pulley of the stent allegedly broke.It was further reported that the stent allegedly failed to deploy.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the covera vascular covered stent products that are cleared in the us.The pro code and 510 k number for the covera vascular covered stent products are identified in d2 and g4.H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the stent delivery system was received for evaluation.The safety lock was found in unlocked position and the grip shells of the delivery system were found disassembled.The stent was found in system and in place.The slide block was found disconnected from the proximal sheath, which is considered being the cause for the reported problems to deploy the covered stent using the deployment wheel.In addition, a short video file was provided with a sequence of an angiography, which is presumably the intended placement site of the sfa with a critical restenosis.The tracking anatomy was reported being reasonably difficult.Device compatible accessories were used.Based on evaluation of the sample the inability of an adhesive joint (slide block/tether/diving sheath) to withstand tension force during deployment is confirmed.A definite root cause could not be determined based upon available information.Labeling review: in reviewing the relevant labeling for this product, the potential issue was found addressed.Based on the instructions for use supplied with this product delivery system specific events that could be associated with clinical complications include but are not limited to failure to deploy and high deployment forces.Regarding correct deployment the instructions for use states: "(.) maintain a stationary hold on the white stability sheath during covered stent deployment (¿).Hold the white stability sheath as close as possible to the introducer without touching the dark brown moving catheter of the distal catheter assembly.Maintain the remainder of the white stability sheath (.) relaxed and avoid tension." regarding preparation the instructions for use states: "pre-dilate the stenosis with a pta balloon catheter of appropriate length and diameter for the lesion to be treated." h10: (expiry date:11/2022).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COVERA PLUS VASCULAR COVERED STENT
Type of Device
VASCULAR COVERED STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key14942928
MDR Text Key301732366
Report Number9681442-2022-00205
Device Sequence Number1
Product Code PFV
UDI-Device Identifier00801741136092
UDI-Public(01)00801741136092
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
P170042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAASLE06100
Device Lot NumberANEY4315
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/13/2022
Initial Date FDA Received07/07/2022
Supplement Dates Manufacturer Received09/22/2022
Supplement Dates FDA Received09/27/2022
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-