Model Number N/A |
Device Problems
Mechanical Problem (1384); Failure to Power Up (1476); Failure to Cut (2587); Calibration Problem (2890)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2022 |
Event Type
malfunction
|
Event Description
|
It was reported that the unit was out of calibration and was cutting too deep.The event occurred during surgery, however, there was no known impact or consequence to the patient.There was no surgical delay.The surgical technique was utilized.No adverse events were reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
(b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was further reported that there was no patient harm.Attempts have been made and no further information has been provided.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d4; g3; g6; h1; h2; h3; h4; h6 the event cannot be confirmed.Review of the most recent repair record determined the motor was not running and the device was out of calibration and the motor, swivel, bearings, shaft, reciprocating arm, control bar, and lever were replaced and resolved the reported issue.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was further reported that after further evaluation, the motor was not running and needs to be replaced.The swivel bearing stack, eccentric shaft, and bearings will all need to be replaced as well as the control bar and shaft.The calibration is off.Attempts have been made and no further information has been provided.
|
|