• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. HIGH FLOW INSUFFLATION UNIT Back to Search Results
Model Number UHI-4
Device Problems Detachment of Device or Device Component (2907); Material Deformation (2976)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2022
Event Type  malfunction  
Manufacturer Narrative
The subject device was returned to an olympus service center for evaluation.During inspection and testing, it was observed the socket of the co2 (carbon dioxide) insufflation connector was deformed.In addition, the left side of the front panel inside the device was damaged, resulting in the insufflation connector being detached from the front panel.The investigation is ongoing; therefore, the root cause of the reported event cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.
 
Event Description
The customer reported the front panel of the subject device was cracked, and the gas cylinder interface was deformed.The device is an olympus loaner device.An olympus field service engineer (fse) further explained that the screw at the front panel connection was broken, and the connection thread was deformed, so the cylinder could not be installed normally.There was no patient or procedure involvement.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the cause of the deformed gas cylinder interface and socket of the co2 gas-injection connector could not be identified.Per the legal manufacturer, the other device defect included in the initial mdr (the screw at the front panel connection was broken) has no potential to cause or contribute to death or serious injury if the malfunctions were to recur.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIGH FLOW INSUFFLATION UNIT
Type of Device
HIGH FLOW INSUFFLATION UNIT
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key14945590
MDR Text Key304427349
Report Number8010047-2022-11482
Device Sequence Number1
Product Code HIF
UDI-Device Identifier04953170324147
UDI-Public04953170324147
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K122180
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 08/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUHI-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2022
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/07/2022
Initial Date FDA Received07/07/2022
Supplement Dates Manufacturer Received07/15/2022
Supplement Dates FDA Received08/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/16/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-