• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. LGN XLPE PS INSERT SZ 3-4 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. LGN XLPE PS INSERT SZ 3-4 15MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 71453249
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2022
Event Type  malfunction  
Event Description
It was reported that, during set up for a tka surgery, when opening a lgn xlpe ps insert sz 3-4 15mm, it was found that it was already open.The procedure was finished with a smith and nephew back up device without surgical delay.No further complications were reported.
 
Manufacturer Narrative
(b)(4).
 
Manufacturer Narrative
H10: h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device's packaging was found to be compromised along the seal, rendering the device inoperative.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the sterilization records revealed the batch was sterilized within normal parameters.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.A contribution of the device to the reported event could be corroborated as the device's packaging was found to be unsealed.Some potential probable causes for this event could include packaging damage in transit or storage.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN XLPE PS INSERT SZ 3-4 15MM
Type of Device
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key14951337
MDR Text Key302492036
Report Number1020279-2022-03259
Device Sequence Number1
Product Code JWH
UDI-Device Identifier00885556255872
UDI-Public00885556255872
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K071071
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71453249
Device Catalogue Number71453249
Device Lot Number21HM02935
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/16/2022
Initial Date FDA Received07/07/2022
Supplement Dates Manufacturer Received08/03/2022
Supplement Dates FDA Received08/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/04/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-