Model Number CDHFA500Q
Device Problems Failure to Capture (1081); No Pacing (3268)
Patient Problem Arrhythmia (1721)
Event Date 06/24/2022
Event Type  Injury  
Event Description
During an in clinic follow-up, a left ventricular capture testing was performed and loss of capture occurred which resulted in the patient experiencing asystole.The device had no backup pulse to prevent the loss of capture and the test was terminated by the healthcare provider to resolve the event.The patient was stable and experienced no adverse consequence due to the asystole.
 
Manufacturer Narrative
The report of lack of ¿back-up¿ safety pacing when running lvcap confirm set-up test when programmed to lv only was confirmed.A software investigation was performed by reviewing session records and/or data logs provided from the customer.The event was a result of a software anomaly where the parameter that ensures delivery of an high output rv pulse when running lvcap confirm was not set correctly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT HF
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key14957812
MDR Text Key295502687
Report Number2017865-2022-13771
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067032010
UDI-Public05415067032010
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2023
Device Model NumberCDHFA500Q
Device Catalogue NumberCDHFA500Q
Device Lot NumberS000080864
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/24/2022
Initial Date FDA Received07/08/2022
Supplement Dates Manufacturer Received11/20/2022
Supplement Dates FDA Received11/07/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/05/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
-
-