Device evaluated by mfr: returned product consisted of an innova self-expanding stent system.The outer sheath, tip, inner sheath, and the remainder of the device were checked for damage.Visual examination revealed the thumbwheel lock had been removed and not returned and the pull grip was partially out.Microscopic examination revealed the stent was partially deployed 1.1 cm from the outer shaft.A kink at the nosecone was confirmed.The flushing luer was not attached nor returned.
|
Reportable based on device analysis completed on 15jun2022.It was reported that the device was difficult to advance.A 8 x 60 x 130 innova was selected for use in the superficial femoral artery (sfa).During insertion, the device was difficult to advance.The procedure was completed with a different device.There were no patient complications.However, device analysis revealed stent premature deployment.
|