Model Number MR850
Device Problems Break (1069); Mechanical Problem (1384)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/28/2022
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The complaint mr850 respiratory humidifier is currently en-route to fisher & paykel healthcare (f&p) for evaluation.We will provide a follow up report upon completion of investigation.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that the power cord of a mr850 respiratory heated humidifier was damaged with exposed copper wire.There was no reported patient involvement.
 
Event Description
A distributor reported on behalf of a healthcare facility in japan that the power cord of a mr850 respiratory heated humidifier was damaged with exposed copper wire.There was no reported patient involvement.
 
Manufacturer Narrative
(b)(4).Method: the complaint mr850 respiratory humidifier was returned to our fisher & paykel healthcare (f&p) service centre in japan where it was inspected by a trained f&p service technician.Our investigation is thus based on the information provided by the f&p service technician.Results: visual inspection revealed that the power cord was damaged with exposed copper wire.Conclusion: we are unable to determine what caused the observed damaged.During production the electrical connections of the earth wires on all mr850 units are 100% tested for electrical continuity.Any product that fails is rejected.All mr850 units are visually inspected for damaged power cords before release for distribution.This suggests that the damage occurred after it had been distributed.The mr850 product technical manual contains a maintenance schedule which instructs the user to conduct annual visual checking, performance and electrical safety testing of the mr850 heater base.In addition, the product technical manual also states that "all servicing procedures should be followed by a humidifier test, and an electrical safety test to ensure proper operation".The mr850 respiratory humidifier is complaint with the following electrical standards: as/nzs 3200.1.0, can/csa 22.2 no.601.01, ul 60601-1, iec 60601-1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRATORY HUMIDIFIER
Type of Device
BTT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine 92618
9494534000
MDR Report Key14973902
MDR Text Key296858273
Report Number9611451-2022-00618
Device Sequence Number1
Product Code BTT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K110019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/07/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/10/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMR850
Device Catalogue NumberMR850
Device Lot Number2100483892
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/14/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-