• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TOSOH HI-TEC, INC. AIA-900; FLUOROMETER, FOR CLINICAL USE Back to Search Results
Model Number AIA-900
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/22/2022
Event Type  malfunction  
Manufacturer Narrative
A field service engineer (fse) was at the customer site to address the reported event.The fse confirmed the reported error by reviewing the error logs and reproduced the error by running a sample.During troubleshooting, the fse found test cups stuck in the waste chute.The fse cleared and cleaned the dirt and removed the test cups from the chute.The fse reran the sample and the error did not reoccur.The fse validated the analyzer by successfully performing quality control (qc) and result was within acceptable range.No further action was required by the fse.The aia-900 analyzer is functioning as expected.A 13-month complaint history review and service history review for similar complaints was performed for serial number: (b)(4) from 22may2021 through aware date 22jun2022.There were no similar complaints identified during the search period.The aia-900 operator's manual under section 12 - flags and error messages states the following: [4153] c.Trans-z home overrun.Cause: the home sensor s062, which is not supposed to be activated after the cup transfer z-axis moves, was activated.No further operation will take place.A mf flag will be attached to the measurement result.Action: please contact tosoh local representative.Check s6062 and also check to see the cause of slipping, and so on, that occurs when pm061 moves to the limit side.The most probable cause of the reported event was due to the waste chute being backed up with dirt.
 
Event Description
A customer reported repeated error message 4153, c trans z home overrun on the aia-900 analyzer.The site rebooted the analyzer and emptied the waste and the error persist.A field service engineer (fse) was dispatched to address the reported event which resulted in a delayed reporting of patient samples for luteinizing hormone (lh ii).There is no indication of any patient intervention or adverse health consequences due to the delay in reporting of patient results.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIA-900
Type of Device
FLUOROMETER, FOR CLINICAL USE
Manufacturer (Section D)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA  7460042
Manufacturer (Section G)
TOSOH HI-TEC, INC.
1-37, fukugawa minami-machi
shunan, yamaguchi 74600 42
JA   7460042
Manufacturer Contact
jorge porras
shiba-koen first building
3-8-2 shiba
minato-ku 10586-23
JA   1058623
6506368314
MDR Report Key14973919
MDR Text Key304377981
Report Number3004529019-2022-00125
Device Sequence Number1
Product Code KHO
UDI-Device Identifier04560189283992
UDI-Public04560189283992
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971103
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAIA-900
Device Catalogue Number022930
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/22/2022
Initial Date FDA Received07/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-