Model Number N/A |
Device Problems
Electrical /Electronic Property Problem (1198); Material Deformation (2976); Naturally Worn (2988); Power Problem (3010)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/29/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Cmp-(b)(4).Foreign country: (b)(6).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Event Description
|
It was reported that the following unit was sent in for maintenance but found in need of repair due to a defective motor, damaged motor, unstable motor and two worn screws.No adverse events were reported as a result of this malfunction.Attempts have been made and no further information has been provided.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The event is confirmed.Review of the most recent repair record determined the unit was found to have unstable motor speed and a damaged motor and screws.The motor and screws were replaced and resolved the reported issue.The unit was manufactured in october 2019 and has not since been returned for preventative maintenance.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; d9; e1; g3; g6; h1; h2 the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Search Alerts/Recalls
|