Model Number 9735665
Device Problems Connection Problem (2900); Application Program Freezes, Becomes Nonfunctional (4031)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/05/2022
Event Type  malfunction  
Event Description
Medtronic received information regarding a navigation system being used outside of a procedure.It was reported that there was some needed troubleshooting for an error message.It was a localizer not connected error message.There was no patient involved.There was some troubleshooting done to try and resolve the issue, there was a self-test performed.The positioning sensor unit (psu) and pcoip client were disconnected and the switch and power supply were connected.The ethernet cable to the poe (power over ethernet) were swapped to a different port on the switch but there was no change.All the devices from the switch were disconnected and reconnected to the pcoip client after power cycling.The pcoip was able to get online, all the devices were reconnected but the psu was still offline.The ethernet from the poe to the camera was bypassed but there was no power to the camera and it was still disconnected.It was determined that the probable cause may have come from the poe injector.
 
Manufacturer Narrative
Other relevant device(s) are: product id: 9735798, no products have been returned to medtronic for analysis.(b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
H3: system service was performed in the field and the power over ethernet (poe) injector was replaced.The system then passed the system checkout and was found to be fully functional.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STEALTHSTATION¿ S8 SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key14976847
MDR Text Key304600311
Report Number1723170-2022-01059
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00643169722187
UDI-Public00643169722187
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/11/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9735665
Device Catalogue Number9735665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/31/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/02/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
"SEE H10...."
-
-