• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TELEFLEX MEDICAL HORIZON TI ML 6/CART 120/BOX; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TELEFLEX MEDICAL HORIZON TI ML 6/CART 120/BOX; CLIP, IMPLANTABLE Back to Search Results
Model Number IPN914927
Device Problem Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  malfunction  
Event Description
The distributor detected in receiving 1 unit with damaged cartridge.
 
Manufacturer Narrative
Qn# (b)(4).The device history review for the product horizon ti ml 6/cart 120/box lot# 73m2100251 investigation did not show issues related to the complaint.
 
Event Description
The distributor detected in receiving 1 unit with damaged cartridge.
 
Manufacturer Narrative
Qn#(b)(4) per dhr the product horizon ti ml 6/cart 120/box lot# 73m2100251 was manufactured on 12/13/2021 a total of (b)(4)pieces.Lot was released on 01/11/2022.Dhr investigation did not show issues related to complaint.The customer provided two photos for analysis; reference pic1_tc# 1900094048 for the customer supplied photos.The complaint of broken package - sterility compromised was able to be confirmed by the photo.The customer returned one sample of 003200 horizon ti ml 6/cart 120/box for investigation.The returned sample was unopened in its original packaging.Visual inspection revealed a crack along one corner of the packaging, including the sterile seal.The reported defect was confirmed.All horizon clips are subjected to 100% inspection for proper packaging, as well as qc inspection based on aql.It is unlikely that this type of damage was present when the sample left the manufacturing facility.Manufacturing was contacted as part of this complaint investigation.Manufacturing stated that the mark/bend present on the sterile sealing was characteristic of a shipping issue.Therefore, the packaging appears to have been damaged during shipping/handling.A corrective action is not required at this time as the packaging appears to have been damaged during shipping/handling.The reported complaint of "broken package - sterility compromised" was confirmed based upon the sample received.Visual inspection revealed a crack along one corner of the packaging, including the sterile seal.All horizon clips are 100% inspected at manufacturing.It is unlikely that this type of damage was present when the sample left the manufacturing facility.Therefore, the packaging appears to have been damaged during shipping/handling after it left the control of teleflex.Teleflex will continue to monitor and trend on complaints of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HORIZON TI ML 6/CART 120/BOX
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
TELEFLEX MEDICAL
morrisville NC
Manufacturer (Section G)
TELEFLEX MEDICAL
rancho el descanso
tecate 21478
MX   21478
Manufacturer Contact
jasmine brown
3015 carrington mill blvd
morrisville, NC 27560
9193614124
MDR Report Key14980160
MDR Text Key304548207
Report Number3003898360-2022-00262
Device Sequence Number1
Product Code FZP
UDI-Device Identifier24026704696370
UDI-Public24026704696370
Combination Product (y/n)N
Reporter Country CodeBR
PMA/PMN Number
K132658
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 06/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN914927
Device Catalogue Number003200
Device Lot Number73M2100251
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 06/22/2022
Initial Date FDA Received07/11/2022
Supplement Dates Manufacturer Received08/08/2022
Supplement Dates FDA Received08/09/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/13/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-