BOSTON SCIENTIFIC CORPORATION EASYTRAK 2 IS-1; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
|
Back to Search Results |
|
Model Number 4542 |
Device Problems
Fracture (1260); High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/21/2022 |
Event Type
Injury
|
Manufacturer Narrative
|
It was noted the lead was disposed of, thus not expected to be returned.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
|
|
Event Description
|
It was reported that during a routine follow-up visit the physician noted the left ventricular (lv) lead impedance measurements were greater than 3000 ohms.Due to a concern over a lead fracture, the physician scheduled a replacement procedure.During the procedure, the lv lead was explanted and successfully replaced.The physician also changed out the entire cardiac resynchronization therapy defibrillator (crt-d) system during this procedure to upgrade to is-4 leads.It was noted that the lv lead was disposed of during the procedure, thus it is not expected to be returned.
|
|