Model Number 4542
Device Problems Fracture (1260); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2022
Event Type  Injury  
Manufacturer Narrative
It was noted the lead was disposed of, thus not expected to be returned.At this time, the product has not been returned.If the product is returned, analysis will be performed and this report would be updated at that time.
 
Event Description
It was reported that during a routine follow-up visit the physician noted the left ventricular (lv) lead impedance measurements were greater than 3000 ohms.Due to a concern over a lead fracture, the physician scheduled a replacement procedure.During the procedure, the lv lead was explanted and successfully replaced.The physician also changed out the entire cardiac resynchronization therapy defibrillator (crt-d) system during this procedure to upgrade to is-4 leads.It was noted that the lv lead was disposed of during the procedure, thus it is not expected to be returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EASYTRAK 2 IS-1
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key14983781
MDR Text Key295673748
Report Number2124215-2022-25140
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526410901
UDI-Public00802526410901
Combination Product (y/n)Y
Reporter Country CodeIT
PMA/PMN Number
P010012/S024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/11/2013
Device Model Number4542
Device Catalogue Number4542
Device Lot Number145970
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 06/21/2022
Initial Date FDA Received07/12/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/11/2011
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-