MAUDE Adverse Event Report: THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC SWITZERLAND GMBH CENTRIMAG 2ND GENERATION PRIMARY CONSOLE; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 102958
Device Problems Mechanical Problem (1384); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/20/2022
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer¿s investigation is completed.
 
Event Description
It was reported that there was an issue with the console.It was thought to be the fuse.It was a power issue.It was to be used on a patient the next day.The hospital¿s biomed department evaluated the console and it was just the one fuse that was blown.The hospital replaced the one fuse and the issue was resolved.
 
Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of the centrimag console having a ¿power issue¿ due to a damaged fuse was not confirmed.The centrimag console (serial number (b)(6)) was not returned for analysis, and no log files were associated with the reported event.Per additional information, the console¿s damaged fuse was replaced at the hospital¿s biomed department, resolving the issue.The root cause of the reported event was unable to be conclusively determined through this analysis.Review of the device history record for the centrimag console, serial number (b)(6), showed the device was manufactured in accordance with manufacturing and qa specifications.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 9.1 ¿changing fuses¿ instructs users on how to properly exchange the console¿s fuses.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
Manufacturer (Section G)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ   CH-8005
Manufacturer Contact
bob fryc
6035 stoneridge drive
pleasanton, CA 94588
7818528204
MDR Report Key14986140
MDR Text Key303822528
Report Number3003306248-2022-11568
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093832
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number102958
Device Lot NumberL06481
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 06/20/2022
Initial Date FDA Received07/12/2022
Supplement Dates Manufacturer Received07/11/2022
Supplement Dates FDA Received07/13/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/31/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-