EIT EMERGING IMPLANT TECHNOLOGIES GMBH T-HANDLE INSERTER; INTERVERTEBRAL FUSION DEVICE WITH BONE GRAFT, LUMBAR
|
Back to Search Results |
|
Model Number PET30101 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/13/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.This report is for an unknown device/unknown lot.Part and lot numbers are unknown; udi number is unknown.Complainant part is not expected to be returned for manufacturer review /investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
It was reported that during a procedure on (b)(6) 2022, one conduit eit plif t-handle inserter had one of the side handles fall off while inserting cage.There was no surgical delay, surgery was completed successfully.This report is for an unk - cage/spacer: conduit eit - tlif.This is report 1 of 1 for (b)(4).
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10: additional narrative: d9, d10, e1.H3, h4, h6: part pet30101, lot e19di1070: supplier: (b)(6) released on november 19, 2019 with no discrepancies.No non-conformance reports (ncrs) were generated during production.No issues were identified during the manufacturing and release of this device that could have contributed to the problem reported by the customer.H3, h6: a product investigation was completed: visual analysis of the returned sample revealed that the pin of the t-handle inserter has broken, the broken fragment was returned.No other issues were observed.A dimensional inspection was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to excessive/unintended forces.The current and manufactured to drawings were reviewed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.¿based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: corrected data: d1, d3, d4, g1, g4.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Investigation summary: a photo of the alleged complaint device was returned to depuy synthes for evaluation.Visual analysis of the photo revealed that the pin of the inserter was broken.No other product problems were identified.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for device.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a manufacturing related potential cause was not suspected,; therefore, no manufacturing record evaluation is required.
|
|