VENUSA DE MEXICO S.A. DE C.V. TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number PN-004 075 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/16/2022 |
Event Type
malfunction
|
Manufacturer Narrative
|
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.The catheter met specifications during temperature testing and electrical testing.However, the device did not meet specifications during a shaft leak test and an irrigation leak test.Further investigation revealed fluid ingress within the catheter shaft proximal to electrode 4, consistent with the failed shaft leak and irrigation leak tests.The pi irrigation tubing was noted to remain attached to metal irrigation tubing at the deformable body and no damage was noted to the irrigation tubing throughout the device.No tears in the shaft material under electrode rings 2-4 were detected, no leaks from within the handle were detected, and no leaks were observed at the fos or between electrodes 1 and 2.The fluid ingress was only noted proximal to electrode 4.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event and observed leak remains unknown.
|
|
Event Description
|
This report is to advise of an event observed during analysis confirming an irrigation leak, making the event reportable.
|
|