Model Number PN-004 075
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/16/2022
Event Type  malfunction  
Manufacturer Narrative
One 75 mm tacticath quartz contact force ablation catheter was received for evaluation.The catheter met specifications during temperature testing and electrical testing.However, the device did not meet specifications during a shaft leak test and an irrigation leak test.Further investigation revealed fluid ingress within the catheter shaft proximal to electrode 4, consistent with the failed shaft leak and irrigation leak tests.The pi irrigation tubing was noted to remain attached to metal irrigation tubing at the deformable body and no damage was noted to the irrigation tubing throughout the device.No tears in the shaft material under electrode rings 2-4 were detected, no leaks from within the handle were detected, and no leaks were observed at the fos or between electrodes 1 and 2.The fluid ingress was only noted proximal to electrode 4.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event and observed leak remains unknown.
 
Event Description
This report is to advise of an event observed during analysis confirming an irrigation leak, making the event reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTICATH¿ QUARTZ CONTACT FORCE ABLATION CATHETER, 75MM
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX  32470
Manufacturer (Section G)
VENUSA DE MEXICO S.A. DE C.V.
calle hertz 1525-6
parque industrial j. bermudez
ciudad juarez, chihuahua 32470
MX   32470
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key14990636
MDR Text Key302022202
Report Number9680001-2022-00030
Device Sequence Number1
Product Code OAE
UDI-Device Identifier07640157990040
UDI-Public07640157990040
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P130026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/12/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/12/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/25/2022
Device Model NumberPN-004 075
Device Catalogue NumberPN-004075
Device Lot Number7526060
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/07/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-