Model Number CI-1400-02H |
Device Problems
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
|
Patient Problems
Failure of Implant (1924); Skin Erosion (2075)
|
Event Date 03/04/2021 |
Event Type
Injury
|
Event Description
|
The recipient was reportedly explanted due to medical reasons.Advanced bionics is in the process of obtaining additional information.When additional information is received, a supplemental report will be submitted.
|
|
Manufacturer Narrative
|
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
The recipient's device was reportedly explanted due to flap necrosis.The recipient was reimplanted with another cochlear device.The recipient is reportedly doing okay.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|
Manufacturer Narrative
|
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed cuts in the silicone overmold on the top and bottom covers, as well as a severed electrode.These are believed to have occurred during revision surgery.The device passed photographic imaging inspection.System lock was verified.The device passed the electrical and mechanical test performed.This device was explanted for medical reasons.The device passed the tests performed.This version of the hires 90k device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
|
|